Tube Cleaning (Cycle Time: 30 minutes)
(Must be done first before running disinfection cycle)
*FILL WATER TANK AND EMPTY SEWAGE TANK PRIOR TO STARTING CYCLE
Install diaper cup holder to side of main body unit adjacent to the control panel.
Remove Detachable Cushion from Diaper Cup. Replace the Diaper Cushion with the appropriate Gender Plastic Cover for cleaning. Push in the posterior part of the Gender Plastic Cover until you hear the ‘click’.
Close the bottom cap of the Tube and fill with 5ml undiluted Medizyme, then close the top cap. Open the Bottom Cap and insert the Tube into the opening in the Gender Plastic Cover for cleaning. Open the TOP Silicon Cap of Disinfection Tube to allow the Medizyme Disinfectant Solution to empty.
Push the menu button on the control panel and select 1-037 or hold the right arrow down for more than 5 seconds to change the mode to Tube Cleaning. Push ‘OK’. Be sure to release the Tube from the Tube Hanger (on the back of Main Body Unit) before use. Once the ‘CLEANING COMPLETE’ message appears, lift the Main Body Unit Cover. Disconnect the Tubing Connector by sliding the Sewage Tank Disconnecting Switch away from the Sewage Tank.
Remove and fill the Water Tank. Remove and empty/clean the Sewage Tank. Replace the Water Tank and Sewage Tank in the Main Body Unit and close cover.